Title :
Modulation of drug delivery rate by hydrogel-incorporating MEMS devices
Author :
Gu, Yuandong ; Bald, Antonio ; Ziaie, Babak ; Siegel, Ronald A.
Author_Institution :
Dept. of Pharm., Minnesota Univ., Minneapolis, MN, USA
fDate :
6/24/1905 12:00:00 AM
Abstract :
We describe two approaches for modulating fluid flow and drug delivery rate in response to external stimuli using hydrogels incorporated in MEMS devices. The first design, a hydrogel-gated flow controller (HFC), consists of two components, a 3-dimensional crosscut structure and a loaded hydrogel. For a temperature-sensitive HFC, temperature cycling between 25 and 40°C results in a flow rate change between 0 and 12 ml/minute, with a 30 second response time. In the second design, a hydrogel-actuated microvalve (HAM) was constructed. In such a device, a hydrogel disc is sandwiched between a porous plate and a flexible silicone rubber membrane. Swelling of the hydrogel produced by diffusion of chemical species through the porous plate, resulting in the deflection of the membrane and closure of the valve intake orifice. A HAM loaded with phenylboronic acid (PBA)-based glucose-sensitive hydrogel was tested. This glucose-sensitive HAM opens and closes in response to changes in glucose concentration and pH. The fastest response time achieved was 16 minutes using a 70 μm thick hydrogel and a 60 μm porous back plate
Keywords :
biosensors; drug delivery systems; flow control; gels; microactuators; microfluidics; microsensors; microvalves; silicone rubber; 16 min; 25 degC; 3-dimensional crosscut structure; 30 s; 40 degC; 60 micron; 70 micron; chemical species; diffusion; drug delivery rate modulation; external stimuli; flexible silicone rubber membrane; flow rate change; fluid flow; glucose concentration; hydrogel disc; hydrogel-actuated microvalve; hydrogel-gated flow controller; hydrogel-incorporating MEMS devices; loaded hydrogel; pH; phenylboronic acid -based glucose-sensitive hydrogel; porous back plate; porous plate; response time; swelling; temperature cycling; temperature-sensitive HFC; valve intake orifice; Biomembranes; Delay; Drug delivery; Fluid flow; Fluid flow control; Hybrid fiber coaxial cables; Microelectromechanical devices; Micromechanical devices; Microvalves; Temperature;
Conference_Titel :
Microtechnologies in Medicine & Biology 2nd Annual International IEEE-EMB Special Topic Conference on
Conference_Location :
Madison, WI
Print_ISBN :
0-7803-7480-0
DOI :
10.1109/MMB.2002.1002356