• DocumentCode
    1600359
  • Title

    Adaptive software for wattmeter calibration

  • Author

    Carminati, Edoardo

  • Author_Institution
    Dipartimento di Electtrotecnica, Politecnico di Milano, Italy
  • fYear
    1991
  • Firstpage
    476
  • Lastpage
    482
  • Abstract
    A methodology is presented which is based on an adaptive software that increases the performances of a measurement system devoted to the characterization of industrial wattmeters working in presence of significant distortions in current and voltage signals and/or low values of cos Φ. The system is a two-channels standard waveform including power outputs (mas. 150 Vrms for the voltage channel and 5 Arms for the current channel). The integrated action of hardware and software resources guarantees a high-level accuracy both in generation of standard waveforms and in sampling and evaluating the power signals. The system provides automatic calibration through feedback path and the correction of systematic errors
  • Keywords
    calibration; computerised instrumentation; wattmeters; adaptive software; automatic calibration; correction; feedback; high-level accuracy; industrial wattmeters; integrated action; standard waveforms; systematic errors; two-channels standard waveform; wattmeter calibration; Calibration; Computer industry; Current measurement; Defense industry; Distortion measurement; Performance evaluation; Software measurement; Software performance; Voltage; Wattmeters;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Instrumentation and Measurement Technology Conference, 1991. IMTC-91. Conference Record., 8th IEEE
  • Conference_Location
    Atlanta, GA
  • Print_ISBN
    0-87942-579-2
  • Type

    conf

  • DOI
    10.1109/IMTC.1991.161639
  • Filename
    161639