DocumentCode
1606164
Title
Sound intensity measurement in focal region of HIFU transducer adopting narrow-pulse method
Author
Changping, Zhu ; Minglei, Shan ; Shichuan, He ; Jia, Zhu ; Ruo, Feng ; Jianyi, Xu ; Qiang, Zhang
Author_Institution
Dept. of Comput. & Inf. Eng., Hohai Univ., Changzhou, China
Volume
3
fYear
2004
Firstpage
1374
Abstract
To meet the requirement for more efficient and convenient method of high-intensity focused ultrasound (HIFU for short) sound field measurement, which is propelled by the development of HIFU therapy, this paper theoretically discusses the feasibility to measure the focal region sound intensity of HIFU transducer by hydrophone method that uses narrow modulated wave as inspiring signal. It is proved through experiments to be a quick, efficient, safe and reliable method for the measurement of HIFU sound intensity.
Keywords
acoustic field; acoustic intensity measurement; hydrophones; ultrasonic transducers; ultrasonic variables measurement; high-intensity focused ultrasound sound intensity measurement; hydrophone method; narrow-pulse method; sound field measurement; transducer; Acoustic measurements; Acoustic transducers; Breast neoplasms; Intensity modulation; Liver neoplasms; Medical treatment; Pressure measurement; Sonar equipment; Ultrasonic imaging; Ultrasonic variables measurement;