DocumentCode :
1614156
Title :
Validation of electron beam position monitoring probes by comparison of dose profile measurements with beam displacement sensing in the microwave range
Author :
Ruf, Marcel ; Schmidt, Lorenz-Peter ; Setzer, Stefan
Author_Institution :
High Freq. Technol., Univ. of Erlangen-Nuremberg, Nuremberg, Germany
fYear :
2011
Firstpage :
1410
Lastpage :
1413
Abstract :
The development of a beam position monitoring and control architecture for compact linear accelerators requires novel sensors which fulfill the tight space restrictions given in such systems. This paper presents simulation and experimental data of capacitive pickup probes with high sensitivity capable of sensing beam deflections far below 1 mm and compact enough to be integrated in accelerating structures. The correlation of beam position monitoring output signals with dose profile measurements is presented indicating that the proposed compact sensors together with dedicated BPM hardware allow reliable electron beam trajectory determination.
Keywords :
dosimetry; electron beams; linear accelerators; microwave spectra; position measurement; trajectory control; BPM hardware; beam displacement sensing; capacitive pickup probes; compact linear accelerators; control architecture; correlation; dose profile measurements; electron beam position monitoring probes; electron beam trajectory; microwave range; sensing beam deflections; Current measurement; Linear accelerators; Monitoring; Particle beams; Position measurement; Probes; Sensitivity; Beam Position Monitoring; Electron Linacs;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Microwave Conference Proceedings (APMC), 2011 Asia-Pacific
Conference_Location :
Melbourne, VIC
Print_ISBN :
978-1-4577-2034-5
Type :
conf
Filename :
6174025
Link To Document :
بازگشت