• DocumentCode
    1617259
  • Title

    Collection efficiency of CdZnTe detectors

  • Author

    Giakos, G.C. ; Vedantham, S. ; Odogba, J.

  • Author_Institution
    Dept. of Biomed. Eng., Akron Univ., OH, USA
  • Volume
    1
  • fYear
    1997
  • Firstpage
    8
  • Abstract
    In this work, the detection of signal charge produced by X-rays incident on resistive CdZnTe semiconductor substrates, at different detector geometries, within the X-ray diagnostic energy range, is investigated. The experimental results suggest that the observed signal-to-noise ratio is dependent upon the choice of the polarizing electrode that is directly exposed to the incident X-ray beam. An efficient charge defection is achieved achieved when the X-ray beam is incident on the negative electrode with the electric field parallel to the direction of the incoming photons
  • Keywords
    II-VI semiconductors; X-ray detection; biomedical equipment; cadmium compounds; diagnostic radiography; semiconductor counters; semiconductor device noise; zinc compounds; CdZnTe; X-ray diagnostic energy range; X-ray produced; collection efficiency; detection of signal charge; digital radiography detectors; efficient charge defection; negative electrode; planar design; polarizing electrodes; resistive II-VI semiconductor substrates; signal-to-noise ratio; solid state ionization detectors; Charge carrier processes; Conductivity; Electrodes; Electron mobility; Geometry; Signal to noise ratio; Substrates; Tellurium; X-ray detection; X-ray detectors;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Instrumentation and Measurement Technology Conference, 1997. IMTC/97. Proceedings. Sensing, Processing, Networking., IEEE
  • Conference_Location
    Ottawa, Ont.
  • ISSN
    1091-5281
  • Print_ISBN
    0-7803-3747-6
  • Type

    conf

  • DOI
    10.1109/IMTC.1997.603906
  • Filename
    603906