Title :
Singles transmission scans performed post-injection for quantitative whole body FDG-PET
Author :
Smith, R.J. ; Benard, F. ; Karp, J.S. ; Muehllehner, G. ; Gualtieri, E.
Author_Institution :
Dept. of Radiol., Pennsylvania Univ., Philadelphia, PA, USA
Abstract :
Post-injection singles transmission scanning has been implemented in the septumless PENN PET 240H scanner (prototype of the GE QUEST). The method uses a 6 mCi point transmission source of 137Cs at the axial center and 37 cm off transaxial center of the camera field of view. Singles transmission scans of 1.8 minutes per bed axial position provide similar scan count densities to 15 minute coincidence transmission scans with a 0.5 mCi 68Ge rod transmission source. Scatter and emission contamination suppression are achieved by applying a narrow 662 keV transmission photopeak energy window. The residual 511 keV emission contamination constitutes a background of uniform spatial distribution. Accurate and low noise attenuation correction is achieved by segmenting the singles transmission images into lung and soft tissue volumes. Established 511 keV gamma ray attenuation coefficients are then applied and these images are forward projected for attenuation correction. Expectation maximisation or OS-EM reconstruction of the transmission and emission images is used to improve image quality. Both the segmentation and OS-EM reconstruction maintain quantitative accuracy in the fully corrected emission images compared to measured coincidence transmission correction. Thus, a clinical protocol involving 40 minutes of emission scans followed by 20 minutes of singles transmission scans allow the 60 cm of the human torso to be fully scanned within 60 minutes. These quantitative whole body FDG PET images may then be used for tumor grading and assessment of tumor response to treatment
Keywords :
error correction; gamma-ray absorption; image reconstruction; image segmentation; lung; maximum likelihood estimation; medical image processing; positron emission tomography; 0.5E-3 Ci; 511 keV; 60 cm; 60 min; 662 keV; 6E-3 Ci; 137Cs point transmission source; Cs; GE QUEST prototype; Ge; OS-EM reconstruction; attenuation correction; camera field of view; clinical protocol; coincidence transmission correction; emission contamination suppression; emission images; expectation maximisation; forward projection; gamma ray attenuation coefficients; human torso; image quality; low noise attenuation correction; lung; post-injection; quantitative accuracy; quantitative whole body FDG-PET; scan count densities; scatter contamination suppression; segmentation; septumless PENN PET 240H scanner; singles transmission scans; soft tissue volumes; transmission images; transmission photopeak energy window; tumor grading; tumor treatment response assessment; uniform spatial distribution; Attenuation; Cameras; Contamination; Image reconstruction; Image segmentation; Lungs; Neoplasms; Positron emission tomography; Prototypes; Scattering;
Conference_Titel :
Nuclear Science Symposium, 1996. Conference Record., 1996 IEEE
Conference_Location :
Anaheim, CA
Print_ISBN :
0-7803-3534-1
DOI :
10.1109/NSSMIC.1996.587916