• DocumentCode
    1622578
  • Title

    E.M.F. drift and inhomogeneity of type K thermocouples

  • Author

    Hamada, Takahiro ; Suyama, Y.

  • Author_Institution
    Tanaka Kikinzoku Kogyo K.K. Isehara Works, Japan
  • Volume
    2
  • fYear
    2004
  • Firstpage
    989
  • Abstract
    Type K thermocouples were exposed at some temperatures in order to estimate their E.M.F. drift and inhomogeneity. About 3/spl deg/C E.M.F. drift was observed at 700/spl deg/C exposure. And over 10/spl deg/C E.M.F. change caused by inhomogeneity was also observed at 800/spl deg/C exposure. It is dangerous for type K thermocouples to calibrate at around 800/spl deg/C.
  • Keywords
    electric potential; thermocouples; 10 C; 3 C; 700 C; 800 C; EMF drift; type K thermocouples;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    SICE 2004 Annual Conference
  • Conference_Location
    Sapporo
  • Print_ISBN
    4-907764-22-7
  • Type

    conf

  • Filename
    1491559