DocumentCode
1625299
Title
The Simple Monitoring method of Trichloroethylene Using a Quartz Crystal Microbalance by the Hybrid type
Author
Noda, Kazutoshi ; Aizawa, H. ; Kurosawa, Shunsuke
Author_Institution
Inst. of Environ. Manage technol., Nat. Inst. of Adv. Ind. Sci. & Technol., Ibaraki
fYear
2006
Firstpage
1656
Lastpage
1659
Abstract
A highly sensitive method for the detection of trichloroethylene (TCE) using a quartz crystal microbalance (QCM) is described. This method (hybrid type QCM) uses lead dioxide and sulfuric acid as oxidizing agents to convert TCE into hydrogen chloride, which in turn reacts directly with a copper electrode on the quartz crystal surface. The results indicate that the sensitivity increases at higher temperatures. Using a calibration curve, it is possible to obtain a TCE gas concentration. This hybrid type QCM method allows for the measurement of very low concentrations of TCE, and has a detection limit comparable to the Japanese Environmental Standard (0.2 mg/m3)
Keywords
calibration; chemical variables measurement; gas sensors; microbalances; organic compounds; Japanese Environmental Standard; QCM; calibration curve; copper electrode; gas concentration; hydrogen chloride; lead dioxide; oxidizing agents; quartz crystal microbalance; sensitive method; sulfuric acid; trichloroethylene monitoring; Atmospheric measurements; Biomembranes; Copper; Electrodes; Frequency; Gas detectors; Hydrogen; Lead compounds; Monitoring; Pollution measurement; Gas sensor; Hybrid; QCM; Trichloroethylene; Volatile organic compounds;
fLanguage
English
Publisher
ieee
Conference_Titel
SICE-ICASE, 2006. International Joint Conference
Conference_Location
Busan
Print_ISBN
89-950038-4-7
Electronic_ISBN
89-950038-5-5
Type
conf
DOI
10.1109/SICE.2006.315586
Filename
4109237
Link To Document