DocumentCode
1634106
Title
A sub-ppm automated 10-to-1 kV measuring system [resistive voltage divider]
Author
Chien, Forest S S ; Su, T.H. ; Chen, I.P. ; Kuo, J.J.
Author_Institution
Center for Meas. Standards, Hsinchu, Taiwan
Volume
1
fYear
1997
Firstpage
350
Abstract
We describe an automated dc 10-to-1 kV measuring system built with simple equipment. This PC-controlled system performs a quick automated self-calibration and sub-ppm accuracy voltage measurement. The bridge method is adopted in its self-calibration procedure. The analysis of the circuits for self-calibration and voltage measurement is included
Keywords
bridge instruments; calibration; voltage dividers; voltage measurement; 10 V to 1 kV; PC-controlled system; automated measuring system; automated self-calibration; bridge method; resistive voltage dividers; sub-ppm accuracy voltage measurement; Automatic voltage control; Bridge circuits; Calibration; Circuit analysis; Electrical resistance measurement; Laboratories; Measurement standards; Performance evaluation; Voltage measurement; Wire;
fLanguage
English