Abstract :
The IEC 61010-1 3rd Edition standard for measurement and control equipment for laboratory use stipulates that a 500VA dielectric withstand tester must be used to test product insulation as part of testing for electrical hazards. This requirement harks back to a time when most withstand tester outputs were unregulated and could cause problems with load regulation. Advancements in withstand tester technology include testers with regulated outputs. This calls into question the need for a 500VA requirement in the IEC 61010-1 3rd Edition standard.
Keywords :
IEC standards; electrical safety; hazards; insulation; load regulation; test equipment; IEC 61010-1 3rd Edition standard; apparent power 500 VA; control equipment; dielectric withstand tester; electrical hazards; hipot testers; load regulation; product insulation test; Dielectrics; Electric breakdown; IEC standards; Insulation; Safety; Testing;