Title :
Real time measurement of intracavitary probe position [cardiology]
Author :
Kun, Stevan ; Peura, Robert A.
Author_Institution :
Dept. of Biomed. Eng., Worcester Polytech. Inst., MA, USA
Abstract :
An experimental system, based on the principles of impedance plethysmography, is tested for continuous real-time computerized determination of the position of a multielectrode intracavitary probe. From the results of the theoretical analysis and mathematical model simulations, tolerances of probe position determination are predicted to be within 1 mm or 5% of the measured distance, whichever is larger. Simulated tank model measurements produced results which indicate that these theoretical predictions are correct
Keywords :
biomedical measurement; cardiology; position measurement; continuous real-time computerized determination; experimental system; impedance plethysmography; intracavitary probe position; mathematical model simulations; simulated tank model measurements; Cardiology; Computational modeling; Impedance; Mathematical model; Plethysmography; Position measurement; Predictive models; Probes; Real time systems; System testing;