• DocumentCode
    1643135
  • Title

    Assessment of iterative image reconstruction techniques for small-animal PET imaging applications

  • Author

    Karali, E. ; Pavlopoulos, S. ; Koutsouris, D.

  • Author_Institution
    Dept. of Electr. & Comput. Eng., Nat. Tech. Univ. of Athens, Athens
  • fYear
    2008
  • Firstpage
    1
  • Lastpage
    6
  • Abstract
    The purpose of this study is to assess the performance of iterative reconstruction methods, using phantom data from a prototype small-animal PET system. The algorithms compared are the simultaneous versions of ART (SART), EM-ML, ISRA WLS and a new iterative algorithm we have introduced under the short name ISWLS. The evaluation study was based on reconstructed image quality, as it is derived from visual inspection, cross-correlation coefficient and CNRs (contrast-to-noise ratios) of specific ROIs (region-of-interest). In general EM-ML and ISRA present similar reconstruction time and minor differences in reconstructed image quality. Slightly superior performances show WLS and SART while ISWLS improves reconstruction resolution at the edges of the field of view.
  • Keywords
    image reconstruction; iterative methods; medical image processing; phantoms; positron emission tomography; ISWLS algorithm; cross correlation coefficient; iterative image reconstruction; phantom; small animal PET imaging; visual inspection; Image quality; Image reconstruction; Imaging phantoms; Inspection; Iterative algorithms; Iterative methods; Positron emission tomography; Prototypes; Reconstruction algorithms; Subspace constraints;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    BioInformatics and BioEngineering, 2008. BIBE 2008. 8th IEEE International Conference on
  • Conference_Location
    Athens
  • Print_ISBN
    978-1-4244-2844-1
  • Electronic_ISBN
    978-1-4244-2845-8
  • Type

    conf

  • DOI
    10.1109/BIBE.2008.4696810
  • Filename
    4696810