Abstract
We present experimental results from studying the impact of waveguide aperture dimensions and misalignment on the calibrated electrical performance of a vector network analyzer from 140 to 325 GHz. It is shown that the calibrated transmission response is not adversely affected by waveguide irregularities or aperture misalignment. In contrast, we observed that the calibrated reflection response is much more sensitive to these mechanical imperfections. Our initial results suggest that waveguide misalignment may have a greater impact on the calibration accuracy of network analyzers operating beyond 140 GHz.
Keywords
millimetre wave devices; network analysers; submillimetre wave devices; waveguides; aperture misalignment; calibrated electrical performance; calibrated transmission response; frequency 140 GHz to 325 GHz; mechanical imperfections; vector network analyzer; waveguide aperture dimensions; waveguide irregularities; waveguide misalignment; Apertures; Performance analysis; calibration; millimeter-wave measurement; waveguide apertures; waveguide misalignment;
fLanguage
English
Publisher
ieee
Conference_Titel
Microwave Measurement Conference, 2009 73rd ARFTG
Conference_Location
Boston, MA
Print_ISBN
978-1-4244-3442-8
Electronic_ISBN
978-1-4244-3443-5
Type
conf
DOI
10.1109/ARFTG.2009.5278062
Filename
5278062
Link To Document