• DocumentCode
    1644982
  • Title

    Orthogonally oriented coils for minimization of cross-coupling in cortical implants

  • Author

    Silay, Kanber Mithat ; Dehollain, Catherine ; Declercq, Michel

  • Author_Institution
    Inst. of Electr. Eng., Ecole Polytech. Fed. de Lausanne, Lausanne
  • fYear
    2008
  • Firstpage
    109
  • Lastpage
    112
  • Abstract
    This article presents the design of the inductive links of remotely powered wireless cortical implants for minimum interference from the power link to the data link. This interference is minimized by changing the orientation of the inductors. Different orientation types are investigated and compared. Orthogonally oriented coils with high interference suppression and better reproducibility has been proposed. From the simulations, 29% power efficiency and 46 dB interference suppression is achieved for these coils. The spiral inductors are fabricated on PCBs and results of preliminary measurements are presented. Moreover, a conceptual cortical implant prototype is proposed.
  • Keywords
    brain; coils; interference suppression; neurophysiology; power inductors; prosthetic power supplies; telemedicine; PCB; cross-coupling minimization; inductive links design; interference suppression; orthogonally oriented coils; remotely powered wireless cortical implants; spiral inductor; Coils; Downlink; Frequency; Implants; Inductors; Interference suppression; Reproducibility of results; Safety; Spirals; Voltage;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Biomedical Circuits and Systems Conference, 2008. BioCAS 2008. IEEE
  • Conference_Location
    Baltimore, MD
  • Print_ISBN
    978-1-4244-2878-6
  • Electronic_ISBN
    978-1-4244-2879-3
  • Type

    conf

  • DOI
    10.1109/BIOCAS.2008.4696886
  • Filename
    4696886