• DocumentCode
    1645561
  • Title

    A CMOS integrated bacterial sensor for rapid detection of Pseudomonas aeruginosa

  • Author

    Nikkhoo, N. ; Mann, C. ; Gulak, P.G. ; Maxwell, K.

  • Author_Institution
    Dept. of Electr. & Comput. Eng., Univ. of Toronto, Toronto, ON
  • fYear
    2008
  • Firstpage
    213
  • Lastpage
    216
  • Abstract
    An integrated bacterial detection chip is implemented in 0.18 mum CMOS technology. The chip has been tested using pyocins as biological detecting elements along with the study of electrical noise generated in an integrated nanowell to detect the presence of two different bacterial clinical isolates of Pseudomonas aeruginosa. The chip successfully identifies the presence of bacterial strains sensitive to the pyocin in less than 10 minutes. The effect of bacterial cell concentration is also presented in the experimental results. The chip consumes 122 muW from 3.3 V supply for two recording channels and occupies an area of 0.48 mm2 in a 0.18 mum standard CMOS process.
  • Keywords
    CMOS image sensors; bioelectric phenomena; biosensors; cellular biophysics; lab-on-a-chip; microorganisms; CMOS integrated bacterial sensor; Pseudomonas aeruginosa; bacterial cell concentration; bacterial strains; bacteriocin; electrical noise; integrated nanowell; lab-on-chip; power 122 muW; pyocins; size 0.18 mum; voltage 3.3 V; Biomembranes; CMOS process; CMOS technology; Capacitive sensors; Cells (biology); Immune system; Integrated circuit noise; Microorganisms; Pathogens; Semiconductor device measurement; Bacterial sensor; Bacteriocin; Integrated circuit; Lab-on-Chip; Pseudomonas aeruginosa; Pyocin;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Biomedical Circuits and Systems Conference, 2008. BioCAS 2008. IEEE
  • Conference_Location
    Baltimore, MD
  • Print_ISBN
    978-1-4244-2878-6
  • Electronic_ISBN
    978-1-4244-2879-3
  • Type

    conf

  • DOI
    10.1109/BIOCAS.2008.4696912
  • Filename
    4696912