Title :
A Robust Algorithm for Faster SAR Measurement in the 30-6000 MHz Frequency Range
Author :
Douglas, M.G. ; Gessner, A.V. ; Chou, C.-K.
Author_Institution :
Motorola Labs., Corp. EME Res. Lab., Fort Lauderdale, FL
Abstract :
An algorithm to quickly estimate the measured specific absorption rate (SAR) from planar scans is presented. This algorithm results in a savings in measurement time of up to 70%. Previously published work from the authors has been extended here to cover the entire frequency range of 30 MHz - 6 GHz specified in draft international standards. SAR data have been collected on real cellular phones to demonstrate that the previous algorithm is accurate. Numerical electromagnetic simulations were conducted in order to extend the frequency range of the algorithm. It is shown that the new algorithm gives similar results to the previous algorithm in the 150 - 2450 MHz frequency range, but it performs much better in the extended frequency range.
Keywords :
bioelectric phenomena; biological effects of microwaves; biomagnetism; cellular radio; dosimetry; cellular phones; fast SAR measurement; frequency 30 MHz to 6000 MHz; numerical electromagnetic simulations; planar scan; specific absorption rate; Area measurement; Electric variables measurement; Electromagnetic measurements; Equations; Frequency measurement; Imaging phantoms; Numerical simulation; Robustness; Specific absorption rate; Time measurement;
Conference_Titel :
Bioinformatics and Biomedical Engineering, 2008. ICBBE 2008. The 2nd International Conference on
Conference_Location :
Shanghai
Print_ISBN :
978-1-4244-1747-6
Electronic_ISBN :
978-1-4244-1748-3
DOI :
10.1109/ICBBE.2008.192