Title :
Integrated optical surface plasmon resonance biosensor for pesticide analysis
Author :
Harris, R.D. ; Luff, B.J. ; Wilkinson, J.S. ; Wilson, R. ; Schiffrin, D.J. ; Piehler, J. ; Brecht, A. ; Abuknesha, R.A. ; Mouvet, C.
Author_Institution :
Optoelectron. Res. Centre, Southampton Univ., UK
fDate :
11/15/1995 12:00:00 AM
Abstract :
We report the design, fabrication and performance of gold coated waveguide surface plasmon resonance biosensors for the detection of the pesticide simazine in aqueous solution. The operation of the sensor was verified against the biotin-avidin biochemical model. The output power of the sensor showed a decrease of 32% on binding a dual layer of biotin-avidin. Calibration of the sensor against standard simazine solutions using a competitive test format was performed and the system was determined to have a detection limit of 0.22 μg/l
Keywords :
biosensors; calibration; chemical sensors; optical planar waveguides; optical sensors; surface plasmons; water pollution measurement; Au; biotin-avidin biochemical model; calibration; coated waveguide biosensors; design; drinking water; environmental sensors; fabrication; integrated optical; output power; performance; pesticide analysis; planar waveguide; simazine in aqueous solution; surface plasmon resonance biosensor; water pollution;
Conference_Titel :
Optical Techniques for Environmental Monitoring, IEE Colloquium on
Conference_Location :
London
DOI :
10.1049/ic:19951120