Optimal QRS detection algorithms
Author
Cohen, Arnon ; Poluta, Mladen
Author_Institution
Electr. & Comput. Eng., Ben Gurion Univ., Israel
fYear
1991
Firstpage
782
Abstract
Performance measures to compare QRS detection algorithms are suggested and evaluated. The measures are used to compare various linear and nonlinear filters and to determine optimal threshold levels. In low error cases the separation divergence is suggested as a performance measure. For single lead systems this measure becomes an F-ratio measure. A database of more than 2700 complexes is used. With this database several types of linear filters and nonlinear operators are compared and optimal threshold levels are determined
Keywords
electrocardiography; signal processing; F-ratio measure; database; linear filters; nonlinear filters; optimal QRS detection algorithms; optimal threshold levels; single lead systems; Band pass filters; Biomedical measurements; Computerized monitoring; Detection algorithms; Electrocardiography; Finite impulse response filter; Nonlinear filters; Signal to noise ratio; Wiener filter; Working environment noise;