Title :
Experimental and Numerical Analysis in Vitro with a Water-Cooled Microwave Ablation Antenna
Author :
Nan, Qun ; Liu, Youjun ; Yi, Zeng
Author_Institution :
Coll. of Life Sci. & Bioeng., Beijing Univ. of Technol., Beijing
Abstract :
This paper gives a comparison of results obtained from numerical simulation and experimental measurements made in vitro pig liver with a water-cooled microwave ablation antenna. An axis-symmetric finite element model was developed to simulate the temperature field based on the pennes bioheat equation, in which the convection heat transfer of the cooling water was involved. The temperature curves and ablation area in the experiments corresponded well with simulated ones in vitro. The results demonstrated that computer-aided simulation of microwave thermal distribution is an accurate and reliable method which can instruct the clinical operation and surgical planning of microwave ablation in practice.
Keywords :
finite element analysis; hyperthermia; liver; microwave antennas; microwave heating; physiological models; surgery; axis-symmetric finite element model; computer-aided simulation; convection heat transfer; in vitro pig liver; microwave thermal distribution; pennes bioheat equation; surgical planning; temperature field; water-cooled microwave ablation antenna; Antenna measurements; Computational modeling; Electromagnetic heating; In vitro; Microwave antennas; Microwave measurements; Microwave theory and techniques; Numerical analysis; Numerical simulation; Temperature;
Conference_Titel :
Bioinformatics and Biomedical Engineering, 2008. ICBBE 2008. The 2nd International Conference on
Conference_Location :
Shanghai
Print_ISBN :
978-1-4244-1747-6
Electronic_ISBN :
978-1-4244-1748-3
DOI :
10.1109/ICBBE.2008.764