Title :
Impedance measurement by means of a high-stability multiphase DDS generator, with the three-voltage method
Author :
Callegaro, Luca ; Galzerano, Gianluca ; Svelto, Cesare
Author_Institution :
Ist. Elettrotecnico Nazionale Galileo Ferraris, Torino, Italy
fDate :
6/24/1905 12:00:00 AM
Abstract :
The system the intercomparison of unlike impedances of IEN, based on the three-voltmeter method and four-terminal pair definition of the standards, has now been provided with a new direct-digital synthesis generator, having multiple outputs, isolation among the channels, and high amplitude and frequency stability. Experimental results in the new and old configuration are shown and compared.
Keywords :
direct digital synthesis; electric impedance measurement; four-terminal pair definition; high-stability multiphase DDS generator; impedance measurement; standards; three-voltmeter method; Bridge circuits; Current measurement; DC generators; Electric variables measurement; Frequency; Impedance measurement; Master-slave; Phase measurement; Stability; Voltage;
Conference_Titel :
Instrumentation and Measurement Technology Conference, 2002. IMTC/2002. Proceedings of the 19th IEEE
Print_ISBN :
0-7803-7218-2
DOI :
10.1109/IMTC.2002.1007125