    A Monte Carlo study of the acceptance to scattered events in a depth encoding PET camera

  • Author

    Moisan, C. ; Tupper, P. ; Rogers, J.G. ; de Jong, J.K.

  • Author_Institution
    TRIUMF, Vancouver, BC, Canada
  • Volume
    3
  • fYear
    1995
  • Firstpage
    1553
  • Abstract
    The authors present a Monte Carlo study of the acceptance to scattered events in a Depth Encoding Large Aperture Camera (DELAC), a hypothetical PET scanner with the capacity to encode the depth-of-interaction (DOI) of incident γ-rays. The simulation is initially validated against the measured scatter fraction of the ECAT-953B scanner. It is then used to assess the response to scattered events in a PET camera made of position encoding blocks of the EXACT HR PLUS type, modified to have DOI resolution through a variation in the photopeak pulse height. The detection efficiency for 511 keV γ-rays, as well as for those that scattered in the object or left only part of their energy in the block, is studied for several combinations of DOI sensitivities and block thicknesses. The scatter fraction predicted by the simulation for DELACs of various ring radii is compared to that of the ECAT-953B as a function of the energy threshold. The results indicate that the poorer discrimination of object scatters with depth sensitive blocks does not lead to a dramatic increase of the scatter fraction
  • Keywords
    Monte Carlo methods; biomedical equipment; cameras; gamma-ray detection; gamma-ray scattering; positron emission tomography; 511 keV; ECAT-953B scanner; EXACT HR PLUS type; Monte Carlo study; depth encoding PET camera; depth-of-interaction; hypothetical PET scanner; incident γ-rays; measured scatter fraction; medical diagnostic imaging; medical instrumentation; nuclear medicine; position encoding blocks; scattered events acceptance; Apertures; Cameras; Encoding; Energy resolution; Gamma ray detection; Gamma ray detectors; Monte Carlo methods; Object detection; Positron emission tomography; Scattering;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium and Medical Imaging Conference Record, 1995., 1995 IEEE
  • Conference_Location
    San Francisco, CA
  • Print_ISBN
    0-7803-3180-X
  • Type

    conf

  • DOI
    10.1109/NSSMIC.1995.500322
  • Filename
    500322