Abstract :
Dual-energy X-ray Computed Tomography (DECT) is a technique which is designed to allow the determination of energy-independent material properties. In this study, results of computer simulation show that energy-dependent systematic errors exist in the values of attenuation coefficient synthesized using the basis material decomposition technique with acrylic and aluminum as the basis materials, especially when iodine (e.g., from radiographic contrast media) is present in the body. The errors are reduced when acrylic and an iodine-water mixture are used as the basis materials. The authors propose a simple model for the calculation of energy-dependent systematic errors using effective energies at the lower and higher energy windows of the X-ray spectrum used in the DECT system. The calculated errors agree well with the errors observed in the simulation. These results suggest that the observed systematic errors are predominantly due to the energy dependence of the basis material coefficients
Keywords :
computerised tomography; digital simulation; measurement errors; H2O; I; X-ray spectrum; basis material coefficients; dual-energy X-ray CT; energy-dependent systematic errors calculation; energy-independent material properties determination; iodine-water mixture; medical diagnostic imaging; simple model; Aluminum; Attenuation; Computational modeling; Computed tomography; Computer errors; Material properties; Physics computing; Radiography; Radiology; X-ray imaging;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference Record, 1995., 1995 IEEE
Conference_Location :
San Francisco, CA
Print_ISBN :
0-7803-3180-X
Type :
conf
DOI :
10.1109/NSSMIC.1995.500330
Filename :
500330
Link To Document :
بازگشت