• DocumentCode
    1680588
  • Title

    A simple determination system for optimal angiographic viewing angles and QCA parameters

  • Author

    Christiaens, J. ; Van de Walle, Rik ; Lemahieu, Ignace

  • Author_Institution
    ELIS-Medisip, Ghent Univ., Belgium
  • Volume
    2
  • fYear
    2001
  • Firstpage
    327
  • Abstract
    In quantitative coronary angiography (QCA), quantification parameters of the coronary arteries are determined using two dimensional (2D) projections of the three dimensional (3D) vessel structures. Unfortunately, the results of this quantification depend on the viewing angles corresponding with the used projections. To deal with this viewpoint dependency, multiple computer systems have been developed to determine the optimal angiographic viewing angles, the angles corresponding with the viewpoints perpendicular to the vessel segments of interest, which are the best possible viewpoints to perform QCA. We developed a system that determines the optimal viewpoints and QCA parameters without creating a reconstruction. Hence the proposed system is very simple and fast. The variation and deviation on results obtained using images of static phantoms are much smaller the inaccuracies due to the movements of the arteries
  • Keywords
    angiocardiography; medical image processing; optimisation; QCA parameters; coronary arteries; medical image processing; optimal viewing angles; quantification parameters; quantitative coronary angiography; Angiography; Arteries; Bifurcation; Image reconstruction; Image segmentation; Lesions; Quantum cellular automata; Two dimensional displays; Visualization; X-ray imaging;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Image Processing, 2001. Proceedings. 2001 International Conference on
  • Conference_Location
    Thessaloniki
  • Print_ISBN
    0-7803-6725-1
  • Type

    conf

  • DOI
    10.1109/ICIP.2001.958494
  • Filename
    958494