  • Author

    Peynsaert, J. ; Achten, E. ; Claeys, E. ; Van de Walle, R. ; de Clercq, D.

  • Author_Institution
    MR Dept., Ghent Univ., Belgium
  • Volume
    1
  • fYear
    1996
  • Firstpage
    134
  • Abstract
    A new type of ergometer for use in an intense magnetic field (MR-environment) was developed in our department. Ergometry tests exercising the calf muscle of volunteers were performed to study the mechanical reliability and electronic accuracy of the totally computer-controlled device. Though different exercise conditions were examined, the variation in the produced cumulative work never exceeded 3%. Exercise information is graphically displayed on-line. Different protocol parameters can be changed interactively to adapt the study to the needs of the investigation. The combination of these ergometric examinations with magnetic resonance spectroscopy measurements ensures a thorough and very accurate in-vivo investigation of the human calf muscular metabolism
  • Keywords
    biomedical NMR; biomedical equipment; computerised monitoring; muscle; signal detection; calf muscle; computer controlled device; electronic accuracy; ergometer signal acquisition system; ergometric examinations; ergometry tests; exercise conditions; exercise information; highly magnetic environment; in-vivo investigation; magnetic resonance spectroscopy measurements; mechanical reliability; muscular metabolism; protocol parameters; Biochemistry; Biological materials; Biomechanics; Hospitals; Information systems; Magnetic fields; Magnetic resonance; Muscles; Protocols; Spectroscopy;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Signal Processing, 1996., 3rd International Conference on
  • Conference_Location
    Beijing
  • Print_ISBN
    0-7803-2912-0
  • Type

    conf

  • DOI
    10.1109/ICSIGP.1996.567064
  • Filename
    567064