    Randoms distributions for triple-head gamma-camera PET systems

  • Author

    Kolthammer, Jeffrey A. ; Matthews, Christopher G.

  • Author_Institution
    Philips Med. Syst., Cleveland, OH, USA
  • Volume
    3
  • fYear
    2001
  • Firstpage
    1185
  • Abstract
    Understanding the effects of random coincidences in PET is important to the analysis of emission images. The noncircular detector structure of triple-head systems gives rise to different randoms distributions than ring tomographs. Using a simulation method, it is found that the shape of the randoms distributions for triple-head systems is more sensitive to the configuration of the detector heads than to the source distribution. The distribution for triple-head systems is spread across the field-of-view, peaking off-center; the dual-head case peaks in the center of the FOV. We also discuss the difference in the axial distribution for 2D and 3D acquisition and rebinning; the introduction of septa and an axial acceptance angle reduces randoms, illustrating an important trade-off between imaging with and without axial septa.
  • Keywords
    positron emission tomography; acceptance angle; axial distribution; field-of-view; random coincidences; randoms distributions; rebinning; septa; triple-head gamma-camera PET systems; Biomedical imaging; Detectors; Electromagnetic scattering; Geometry; Image analysis; Image reconstruction; Magnetic heads; Medical simulation; Positron emission tomography; Shape;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2001 IEEE
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-7324-3
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2001.1008548
  • Filename
    1008548