DocumentCode
1705006
Title
Validation of the SPECT features of a simulation system for emission tomography
Author
Harrison, Robert L. ; Laymon, Charles M. ; Vannoy, Steven D. ; Lewellen, Thomas K.
Author_Institution
Dept. of Radiol., Washington Univ., Seattle, WA, USA
Volume
3
fYear
2001
Firstpage
1363
Abstract
SimSET (a Simulation System for Emission Tomography) is a widely used simulation package in the nuclear medicine research community. In this paper we present validation measurements for the SPECT-related modules of the package. Decay generation and tracking through the object were checked for correct distribution of starting angle and weight, and correct proportion and spatial distribution of unscattered, Compton and Rayleigh scattered photons. We tested that the collimator module, which models the collimator using geometric transfer functions, implements the underlying equations and positions the collimator correctly. The binning of detected photons into histograms was checked for accuracy. Finally, a comparison of simulated data and experimental data is given.
Keywords
digital simulation; medical computing; single photon emission computed tomography; Compton scattered photons; Rayleigh scattered photons; SPECT-related modules; SimSET; Simulation System for Emission Tomography; decay generation; simulation package; Collimators; Medical simulation; Nuclear medicine; Nuclear power generation; Packaging; Particle scattering; Rayleigh scattering; Single photon emission computed tomography; Solid modeling; Testing;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium Conference Record, 2001 IEEE
ISSN
1082-3654
Print_ISBN
0-7803-7324-3
Type
conf
DOI
10.1109/NSSMIC.2001.1008590
Filename
1008590
Link To Document