Title :
Normalisation of listmode data with application to HiDAC data
Author :
Thielemans, K. ; Morel, C. ; Jacobson, M.W. ; Kaempf, J.H. ; Mustafovic, S.
Author_Institution :
Imaging Res. Solutions Ltd., Hammersmith Hosp., London, UK
Abstract :
Many PET scanners nowadays have the possibility to record event-by-event information, known as listmode data. This has the advantage to keep the data in the highest possible resolution (both temporal and spatial). In most cases, listmode data are then binned into sinogram format before reconstruction. In this paper, we discuss at which stage normalisation factors should be introduced. It is shown that noise is greatly reduced by performing the normalisation after the binning. We also briefly discuss the consequences for strictly listmode-based algorithms. We illustrate this with acquired and simulated data for the quad HiDAC camera.
Keywords :
positron emission tomography; HiDAC camera; PET scanners; listmode data; noise; normalisation factors; sinogram; Animals; Cameras; Cyclotrons; Detectors; Educational institutions; Jacobian matrices; Noise reduction; Positron emission tomography; Spatial resolution; Statistical analysis;
Conference_Titel :
Nuclear Science Symposium Conference Record, 2001 IEEE
Print_ISBN :
0-7803-7324-3
DOI :
10.1109/NSSMIC.2001.1008635