Method for measuring performance of fixation techniques for hamstring tendon graft replacement of the anterior cruciate ligament
Author :
Wheeler, M.W. ; Nissen, C.W. ; Adams, D.J.
Author_Institution :
Biomed. Eng. Dept., Connecticut Univ., Storrs, CT, USA
Abstract :
The purpose of this study is to evaluate the dynamic mechanical performance of fixation techniques used in hamstring tendon reconstruction of the anterior cruciate ligament. Testing is done in-vitro using polyurethane bone-analog foam to maintain a uniform model of the tibial bone. The fixation constructs are evaluated according to cyclic creep/slip performance, failure strength, and stiffness by implementing a new method of individually loading the four strands of hamstring tendon.
Keywords :
biomechanics; bone; creep; elongation; orthopaedics; patient rehabilitation; polymers; slip; surgery; anterior cruciate ligament; cyclic creep/slip performance; dynamic mechanical performance; failure strength; fixation constructs; fixation techniques; hamstring tendon four strands; hamstring tendon graft replacement; in-vitro testing; load-elongation profiles; polyurethane bone-analog foam; stiffness; tibial bone; uniform model; Bones; Fasteners; Interference; Joints; Ligaments; Maintenance engineering; Orthopedic surgery; Surges; Tendons; Testing;