DocumentCode :
1717658
Title :
Effect of photomultiplier gain-drift and radiation exposure on 2D-map decoding of detector arrays used in positron emission tomography
Author :
Uribe, Jorge ; Li, Hongdi ; Baghaei, Hossain ; Aykac, Mehmet ; Wang, Yu ; Liu, Yaqiang ; Xing, Tao ; Wong, Wai-Hoi
Author_Institution :
M.D. Anderson Cancer Center, Texas Univ., Houston, TX, USA
Volume :
4
fYear :
2001
fDate :
6/23/1905 12:00:00 AM
Firstpage :
1960
Lastpage :
1964
Abstract :
Arrays of scintillation crystals optically coupled to four single-anode photomultipliers (PMT) are commonly used in PET cameras. The accuracy of the crystal-position decoding relies on the relative gain stability of all 4 PMT over time. The drift in the gain of one PMT relative to the other three distorts the two dimensional (2-D) crystal-decoding map and causes wrong crystal assignment of a gamma hit. We have studied gain stability and its effect on the 2-D decoding map using 4 Philips XP1911 PMT (19 mm diameter) coupled in the quadrant-sharing mode (PQS) to a position-sensitive BGO array (7×7) over 100-day period. The effect on PMT gain due to radiation exposure of the block was also studied. Using 662 keV gamma rays from a Cs-137 source and the collimated tight of a blue LED coupled to the block we measured the individual PMT gain drift, total composite-energy pulse heights and 2-D decoding maps. Some of the drifts were temperature dependent. The measurements were also corrected for temperature variations of the technique. It was found that throughout the duration of the experiment all four PMT presented similar and continuous gain shifts averaging an 11% drop after 100 days
Keywords :
gamma-ray detection; photomultipliers; position sensitive particle detectors; positron emission tomography; solid scintillation detectors; 100 day; 19 mm; 2D-map decoding; 662 keV; 137Cs; Bi4Ge3O12; Cs; PET; PMT; continuous gain shifts; crystal-position decoding; detector arrays; gamma rays; photomultiplier gain-drift; position-sensitive array; positron emission tomography; quadrant-sharing mode; radiation exposure; relative gain stability; scintillation crystals; single-anode photomultipliers; Cameras; Crystals; Decoding; Optical arrays; Optical coupling; Optical distortion; Photomultipliers; Positron emission tomography; Pulse measurements; Stability;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium Conference Record, 2001 IEEE
Conference_Location :
San Diego, CA
ISSN :
1082-3654
Print_ISBN :
0-7803-7324-3
Type :
conf
DOI :
10.1109/NSSMIC.2001.1009208
Filename :
1009208
Link To Document :