Title : 
A Monte-Carlo simulation study on coarse septa for scatter correction in 3D PET
         
        
            Author : 
Hasegawa, Tomoyuki ; Tanaka, Eiichi ; Yamashita, Takaji ; Watanabe, Mitsuo ; Yamaya, Taiga ; Murayarna, H.
         
        
            Author_Institution : 
Sch. of Allied Health Sci., Kitasato Univ., Sagamihara, Japan
         
        
        
        
            fDate : 
6/23/1905 12:00:00 AM
         
        
        
        
            Abstract : 
This report shows the usefulness of coarse septa in positron emission tomography by Monte-Carlo simulation. It was shown that coarse septa result in higher sensitivity than two-dimensional mode with conventional full-septa and lower scatter fraction than three-dimensional mode without septa. In addition, we propose a new scatter correction method using coarse septa, where scatter components are estimated from coincidence data for those line-of-responses that are intercepted by coarse septa. This scatter correction potentially has an ability to cope with scatter from radioactivity outside the field-of-view
         
        
            Keywords : 
Monte Carlo methods; medical image processing; positron emission tomography; Monte-Carlo simulation; coarse septa; coincidence data; line-of-responses; positron emission tomography; scatter correction method; scatter fraction; Costs; Crystals; Electromagnetic scattering; Energy measurement; Energy resolution; Imaging phantoms; Light scattering; Particle scattering; Photonics; Positron emission tomography;
         
        
        
        
            Conference_Titel : 
Nuclear Science Symposium Conference Record, 2001 IEEE
         
        
            Conference_Location : 
San Diego, CA
         
        
        
            Print_ISBN : 
0-7803-7324-3
         
        
        
            DOI : 
10.1109/NSSMIC.2001.1009240