A Monte-Carlo simulation study on coarse septa for scatter correction in 3D PET
Author
Hasegawa, Tomoyuki ; Tanaka, Eiichi ; Yamashita, Takaji ; Watanabe, Mitsuo ; Yamaya, Taiga ; Murayarna, H.
Author_Institution
Sch. of Allied Health Sci., Kitasato Univ., Sagamihara, Japan
Volume
4
fYear
2001
fDate
6/23/1905 12:00:00 AM
Firstpage
2110
Lastpage
2114
Abstract
This report shows the usefulness of coarse septa in positron emission tomography by Monte-Carlo simulation. It was shown that coarse septa result in higher sensitivity than two-dimensional mode with conventional full-septa and lower scatter fraction than three-dimensional mode without septa. In addition, we propose a new scatter correction method using coarse septa, where scatter components are estimated from coincidence data for those line-of-responses that are intercepted by coarse septa. This scatter correction potentially has an ability to cope with scatter from radioactivity outside the field-of-view
Keywords
Monte Carlo methods; medical image processing; positron emission tomography; Monte-Carlo simulation; coarse septa; coincidence data; line-of-responses; positron emission tomography; scatter correction method; scatter fraction; Costs; Crystals; Electromagnetic scattering; Energy measurement; Energy resolution; Imaging phantoms; Light scattering; Particle scattering; Photonics; Positron emission tomography;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium Conference Record, 2001 IEEE
Conference_Location
San Diego, CA
ISSN
1082-3654
Print_ISBN
0-7803-7324-3
Type
conf
DOI
10.1109/NSSMIC.2001.1009240
Filename
1009240
Link To Document