    A Monte-Carlo simulation study on coarse septa for scatter correction in 3D PET

  • Author

    Hasegawa, Tomoyuki ; Tanaka, Eiichi ; Yamashita, Takaji ; Watanabe, Mitsuo ; Yamaya, Taiga ; Murayarna, H.

  • Author_Institution
    Sch. of Allied Health Sci., Kitasato Univ., Sagamihara, Japan
  • Volume
    4
  • fYear
    2001
  • fDate
    6/23/1905 12:00:00 AM
  • Firstpage
    2110
  • Lastpage
    2114
  • Abstract
    This report shows the usefulness of coarse septa in positron emission tomography by Monte-Carlo simulation. It was shown that coarse septa result in higher sensitivity than two-dimensional mode with conventional full-septa and lower scatter fraction than three-dimensional mode without septa. In addition, we propose a new scatter correction method using coarse septa, where scatter components are estimated from coincidence data for those line-of-responses that are intercepted by coarse septa. This scatter correction potentially has an ability to cope with scatter from radioactivity outside the field-of-view
  • Keywords
    Monte Carlo methods; medical image processing; positron emission tomography; Monte-Carlo simulation; coarse septa; coincidence data; line-of-responses; positron emission tomography; scatter correction method; scatter fraction; Costs; Crystals; Electromagnetic scattering; Energy measurement; Energy resolution; Imaging phantoms; Light scattering; Particle scattering; Photonics; Positron emission tomography;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2001 IEEE
  • Conference_Location
    San Diego, CA
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-7324-3
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2001.1009240
  • Filename
    1009240