We propose a novel approach to intraocular pressure (IOP) measurement allowing minimal invasive continuous monitoring over prolonged periods, regardless of patient position and activities. The key element of this innovative measurement method is a soft contact lens with a microfabricated strain gage embedded allowing the measurement of the cornea curvature deformation due to IOP changes. Tests on a eye-simulating device have shown that the sensor is capable of monitoring continuously a deformation of cornea radius inferior to 0.0055 mm, which corresponds to a sensitivity of better than 2 mmHg IOP, according to an in vitro and an in vivo studies about the correlation between IOP and corneal curvature in human beings.
Keywords
contact lenses; eye; micromechanical devices; strain gauges; MEMS strain gage; cornea curvature deformation; eye-simulating device; human beings; intraocular pressure monitoring; minimal invasive; patient position; soft contact lens; Capacitive sensors; Cornea; In vitro; Lenses; Micromechanical devices; Patient monitoring; Position measurement; Pressure measurement; Strain measurement; Testing;
fLanguage
English
Publisher
ieee
Conference_Titel
TRANSDUCERS, Solid-State Sensors, Actuators and Microsystems, 12th International Conference on, 2003
Conference_Location
Boston, MA, USA
Print_ISBN
0-7803-7731-1
Type
conf
DOI
10.1109/SENSOR.2003.1216947
Filename
1216947
Link To Document