• DocumentCode
    1726367
  • Title

    A soft contact lens with a MEMS strain gage embedded for intraocular pressure monitoring

  • Author

    Leonardi, M. ; Leuenberger, P. ; Bertrand, D. ; Bertsch, A. ; Renaud, P.

  • Author_Institution
    Microsyst. Lab., Ecole Polytech. Fed. de Lausanne, Switzerland
  • Volume
    2
  • fYear
    2003
  • Firstpage
    1043
  • Abstract
    We propose a novel approach to intraocular pressure (IOP) measurement allowing minimal invasive continuous monitoring over prolonged periods, regardless of patient position and activities. The key element of this innovative measurement method is a soft contact lens with a microfabricated strain gage embedded allowing the measurement of the cornea curvature deformation due to IOP changes. Tests on a eye-simulating device have shown that the sensor is capable of monitoring continuously a deformation of cornea radius inferior to 0.0055 mm, which corresponds to a sensitivity of better than 2 mmHg IOP, according to an in vitro and an in vivo studies about the correlation between IOP and corneal curvature in human beings.
  • Keywords
    contact lenses; eye; micromechanical devices; strain gauges; MEMS strain gage; cornea curvature deformation; eye-simulating device; human beings; intraocular pressure monitoring; minimal invasive; patient position; soft contact lens; Capacitive sensors; Cornea; In vitro; Lenses; Micromechanical devices; Patient monitoring; Position measurement; Pressure measurement; Strain measurement; Testing;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    TRANSDUCERS, Solid-State Sensors, Actuators and Microsystems, 12th International Conference on, 2003
  • Conference_Location
    Boston, MA, USA
  • Print_ISBN
    0-7803-7731-1
  • Type

    conf

  • DOI
    10.1109/SENSOR.2003.1216947
  • Filename
    1216947