• DocumentCode
    1728585
  • Title

    An adaptive objective for optical motion correction in MRI

  • Author

    Schneider, F. ; Draheim, J. ; Burger, T. ; Maclaren, J. ; Herbst, M. ; Zaitsev, M. ; Bammer, R. ; Wallrabe, U.

  • Author_Institution
    Dept. of Microsyst. Eng., Univ. of Freiburg, Freiburg, Germany
  • fYear
    2010
  • Firstpage
    189
  • Lastpage
    190
  • Abstract
    We present an adaptive membrane lens that is compatible with a magnetic resonance (MR) imaging system, as a part of an objective for in-bore optical motion tracking and correction. Compatibility tests with the high magnetic fields of a 3 T scanner and the shielding of the adaptive lens are discussed. For tracking of a marker an adaptive wide-angle objective has been designed and characterised. The measured mean distortions of the objective are between -2.05 and -2.6 %. The tracking system was verified in a MR scanner model.
  • Keywords
    adaptive optics; biomedical MRI; lenses; MRI; adaptive membrane lens; adaptive objective; compatibility tests; magnetic resonance imaging; optical motion correction; shielding; Adaptive optics; Distortion measurement; Head; Lenses; Optical distortion; Target tracking;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Optical MEMS and Nanophotonics (OPT MEMS), 2010 International Conference on
  • Conference_Location
    Sapporo
  • Print_ISBN
    978-1-4244-8926-8
  • Electronic_ISBN
    978-1-4244-8925-1
  • Type

    conf

  • DOI
    10.1109/OMEMS.2010.5672120
  • Filename
    5672120