Abstract :
The length of calibration intervals or measurement instrumentations can be determined by means of several techniques. In this paper, three different methods are compared for the establishment of optimal calibration interval of atomic clocks. The first one is based on a stochastic model and provides the estimation of the calibration interval also in the transient situation, while the others, attain to the class of the so-called reactive methods, which determine the value of the optimal calibration interval on the basis of the last calibration outcomes. Algorithms have been applied to experimental data and obtained results have been compared in order to determine the most effective technique. Since the analyzed reactive methods presents a presents a large transient time, a new algorithm is proposed and applied to the available data.
Keywords :
atomic clocks; calibration; stochastic processes; atomic clocks; calibration interval estimation; calibration outcomes; measurement instrumentations; optimal calibration intervals; reactive methods; stochastic model; transient time; Atomic clocks; Calibration; Costs; Instruments; Laboratories; Maintenance; Stochastic processes; Telecommunications; Testing; Time measurement;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Instrumentation and Measurement Technology Conference, 2004. IMTC 04. Proceedings of the 21st IEEE
ISSN :
1091-5281
Print_ISBN :
0-7803-8248-X
Type :
conf
DOI :
10.1109/IMTC.2004.1351006
Filename :
1351006
Link To Document :
بازگشت