• DocumentCode
    1737112
  • Title

    An in-plane microneedles used for sampling and glucose analysis

  • Author

    Liu, R. ; Wang, X.H. ; Tang, F. ; Feng, Y.Y. ; Zhou, Z.Y.

  • Author_Institution
    Dept. of Precision Instruments & Mechanology, Tsinghua Univ., Beijing, China
  • Volume
    2
  • fYear
    2005
  • Firstpage
    1517
  • Abstract
    We report here the design, fabrication and experiments of a hollow in-plane microneedle, with which a biosensor is integrated to extract interstitial fluid from skin for glucose measurement. The design of hollow in-plane microneedles leads to a simple fabrication process. Different from common methods for immobilizing enzyme, we use capillary force to immobilize enzyme on the surface of microelectrodes. The experimental results showed a good linearity for the characteristic curve of glucose concentration, when it varies from 0 to 500mg/dl with respect to current signal value. The sensitivity of 0.4238nA was obtained.
  • Keywords
    biomedical measurement; biosensors; enzymes; microelectrodes; microfluidics; sugar; biosensor; enzyme immobilization; glucose analysis; glucose concentration; glucose measurement; hollow in-plane microneedle; interstitial fluid; microelectrode surface; Biochemical analysis; Biochemistry; Biosensors; Diabetes; Electrodes; Fabrication; Gold; Microelectrodes; Sampling methods; Sugar;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Solid-State Sensors, Actuators and Microsystems, 2005. Digest of Technical Papers. TRANSDUCERS '05. The 13th International Conference on
  • Print_ISBN
    0-7803-8994-8
  • Type

    conf

  • DOI
    10.1109/SENSOR.2005.1497372
  • Filename
    1497372