Simultaneous measurement of percent carboxyhemoglobin and functional oxygen saturation
Author :
Lee, Jason C S ; Cheung, Peter W. ; Marble, David R. ; Kenny, Margaret A. ; Landicho, Del
Author_Institution :
Microsensor Res. Lab., Washington Univ., Seattle, WA, USA
fYear :
1989
Firstpage :
1092
Abstract :
To account for the effect of carboxyhemoglobin on pulse oximeters, the feasibility of simultaneously measuring the functional oxygen saturation and percent carboxyhemoglobin using the pulse oximetry technique was studied. A theoretical model was established, and an in-vitro experiment was carried out. Both the theoretical and experimental results show that the simultaneous measurement is feasible. It was found that the use of 630-nm light gives very good measurement accuracy
Keywords :
biological techniques and instruments; proteins; 630 nm; measurement accuracy; percent carboxyhemoglobin; pulse oximeters; theoretical model; Blood; Hospitals; In vitro; Microsensors; Optical modulation; Oxygen; Pulse amplifiers; Pulse measurements; Table lookup; Wavelength measurement;