• DocumentCode
    1737132
  • Title

    Simultaneous measurement of percent carboxyhemoglobin and functional oxygen saturation

  • Author

    Lee, Jason C S ; Cheung, Peter W. ; Marble, David R. ; Kenny, Margaret A. ; Landicho, Del

  • Author_Institution
    Microsensor Res. Lab., Washington Univ., Seattle, WA, USA
  • fYear
    1989
  • Firstpage
    1092
  • Abstract
    To account for the effect of carboxyhemoglobin on pulse oximeters, the feasibility of simultaneously measuring the functional oxygen saturation and percent carboxyhemoglobin using the pulse oximetry technique was studied. A theoretical model was established, and an in-vitro experiment was carried out. Both the theoretical and experimental results show that the simultaneous measurement is feasible. It was found that the use of 630-nm light gives very good measurement accuracy
  • Keywords
    biological techniques and instruments; proteins; 630 nm; measurement accuracy; percent carboxyhemoglobin; pulse oximeters; theoretical model; Blood; Hospitals; In vitro; Microsensors; Optical modulation; Oxygen; Pulse amplifiers; Pulse measurements; Table lookup; Wavelength measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 1989. Images of the Twenty-First Century., Proceedings of the Annual International Conference of the IEEE Engineering in
  • Conference_Location
    Seattle, WA
  • Type

    conf

  • DOI
    10.1109/IEMBS.1989.96101
  • Filename
    96101