DocumentCode
1740632
Title
Re-evaluation of Compton scattering method in characterization of radiation therapy beams
Author
Li, Texin ; Kim, Siyong ; Liu, Chihray ; Palta, Jatinder ; Tome, Wolfgang ; Ferrari, Angelo ; Park, Joon ; Anghaie, S.
Author_Institution
Dept. of Nucl. Eng., Florida Univ., Gainesville, FL, USA
Volume
2
fYear
2000
fDate
2000
Firstpage
1082
Abstract
To explore the possibility of using Compton scattering method to characterize the energy spectra of clinical beams. This method is experimentally tested on a Co60 radiation therapy machine and independently verified using Monte Carlo (MCNP4B) simulation. Measured spectrum and MCNP simulated spectrum including detector response for 90° scattered beam show good agreement
Keywords
Compton effect; Monte Carlo methods; dosimetry; gamma-ray detection; gamma-ray scattering; radiation therapy; Co60 radiation therapy machine; Compton scattering method; MCNP simulated spectrum; Monte Carlo simulation; clinical beams; detector response; energy spectra; radiation therapy beams characterisation; Biomedical applications of radiation; Charge carrier processes; Electromagnetic scattering; Energy measurement; Geometry; Monte Carlo methods; Particle scattering; Pulse measurements; Radiation detectors; Solids;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 2000. Proceedings of the 22nd Annual International Conference of the IEEE
Conference_Location
Chicago, IL
ISSN
1094-687X
Print_ISBN
0-7803-6465-1
Type
conf
DOI
10.1109/IEMBS.2000.897915
Filename
897915
Link To Document