• DocumentCode
    1740660
  • Title

    A quality assurance program for radiotherapy centers in Korea

  • Author

    Kim, G. ; Oh, H. ; Pyun, W. ; Lee, H.

  • Volume
    2
  • fYear
    2000
  • fDate
    2000
  • Firstpage
    1246
  • Abstract
    Recognizing the importance of quality assurance in radiotherapy and the need to make access to radiation standards traceable to the international measurement system to every radiotherapy center, the KFDA, as a national secondary standard dosimetry laboratory, has started a quality audit program in 1999. This program was initiated by tele-survey to all radiotherapy centers regarding general information about their radiotherapists, medical physicists, type of equipment, dosimeters, etc. This provided the KFDA with a data file and led to links between the KFDA and the clinics. A quality assurance network was set up and on-site dosimetry reviews were arranged according to a suitable timetable. Usually, the outputs of radiotherapy units at given conditions were measured with Farmer-type ionization chambers and compared with corresponding values quoted by the radiotherapy centers
  • Keywords
    calibration; dosimeters; dosimetry; ionisation chambers; quality control; radiation therapy; Farmer-type ionization chambers; Korean radiotherapy centers; calibration; dosimeters; on-site dosimetry reviews; quality assurance program; radiation standards; secondary standard dosimetry laboratory; Calibration; Dosimetry; Imaging phantoms; Ionization chambers; Laboratories; Linear accelerators; Measurement standards; Medical treatment; Quality assurance; Uncertainty;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 2000. Proceedings of the 22nd Annual International Conference of the IEEE
  • Conference_Location
    Chicago, IL
  • ISSN
    1094-687X
  • Print_ISBN
    0-7803-6465-1
  • Type

    conf

  • DOI
    10.1109/IEMBS.2000.897958
  • Filename
    897958