Title :
Improvement of target and critical structure doses in Corvus IMRT treatment plans
Author :
Sethi, Anil ; Leybovich, Leonid ; Dogan, Nesrin ; Krasin, Matthew ; Emami, Bahman
Author_Institution :
Dept. Radiat. Oncology, Loyola Univ., Maywod, IL, USA
Abstract :
The NOMOS MIMIC unit delivers radiation in a tomographic fashion, one arc at a time. Each arc radiates two slices. The slice thickness is determined by the beamlet width: 0.85 cm in the 1 cm MIMIC mode and 1.7 cm in the 2 cm mode. The resolution of beam intensity modulation in the direction of couch motion is defined by the beamlet width. A systematic study of targets of various lengths showed that the dose coverage is a cyclical function of target length. The target dose heterogeneity index (h=Dmax./Dmin.) periodically varied from 1.1 to 1.3 as target length was increased, The period of this function was approximately equal to the beamlet width. Corvus positions treatment arcs symmetrically with respect to the target length. Thus for certain length targets, the ends of target may fall in the beamlet penumbra region thereby reducing target minimum dose. In addition, when a critical structure is located adjacent to a target, the same beamlet bank may irradiate both target and critical structure. This leads to an overdosing of the critical structure and an underdosing of the target. However, small modifications in the target representation (a slight increase in PTV or the addition of a distant pseudo target) can be made to cause a shift in are positions. When the treatment arcs are shifted such that the target is positioned outside the penumbra region and no beamlet bank radiates both target and critical structure, a better target dose distribution and sparing of critical structure is achieved
Keywords :
computerised tomography; dosimetry; intensity modulation; planning; radiation therapy; 0.85 cm; 1.7 cm; NOMOS MIMIC unit; beam intensity modulation; beamlet bank; beamlet penumbra region; beamlet width; corvus IMRT treatment plans; couch motion is; critical structure; critical structure doses; cyclical function; distant pseudo target; dose coverage; overdosing; resolution; slice thickness; target dose heterogeneity index; target length; target minimum dose; target representation; tomographic fashion; treatment arcs; Computed tomography; Cost function; Imaging phantoms; Intensity modulation; Medical treatment; Oncology; Process planning; Radiography; Size control;
Conference_Titel :
Engineering in Medicine and Biology Society, 2000. Proceedings of the 22nd Annual International Conference of the IEEE
Conference_Location :
Chicago, IL
Print_ISBN :
0-7803-6465-1
DOI :
10.1109/IEMBS.2000.898037