DocumentCode :
1741149
Title :
Important issues regarding diode performance in radiation therapy applications
Author :
Shi, Jie ; Simon, William E. ; Ding, Li ; Saini, Daljit
Author_Institution :
Sun Nucl. Corp., Melbourne, FL, USA
Volume :
3
fYear :
2000
fDate :
2000
Firstpage :
1710
Abstract :
Diode detectors have found wide applications in radiation therapy due to their high spatial resolution, high sensitivity and ruggedness. Some of their limitations have been discussed in publications, such as sensitivity degradation with radiation, SSD and field size dependence. This paper addresses some important issues that may not be well recognized-diode impedance, directional dependence, accumulated dose effects
Keywords :
biomedical equipment; dosimetry; radiation therapy; silicon radiation detectors; accumulated dose effects; diode detectors; diode impedance; diode performance; directional dependence; high sensitivity; high spatial resolution; radiation therapy; ruggedness; Biomedical applications of radiation; Diodes; Envelope detectors; Imaging phantoms; Impedance; Leakage current; Particle beams; Temperature dependence; Testing; Voltage;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society, 2000. Proceedings of the 22nd Annual International Conference of the IEEE
Conference_Location :
Chicago, IL
ISSN :
1094-687X
Print_ISBN :
0-7803-6465-1
Type :
conf
DOI :
10.1109/IEMBS.2000.900410
Filename :
900410
Link To Document :