Title :
Dosimetric modelling of a scanned pencil beam apparatus for evaluation of cobalt-60 tomotherapy
Author :
Kerr, Andrew ; Salomons, Greg J. ; Schreiner, L. John
Author_Institution :
Kingston Regional Cancer Centre, Ont., Canada
Abstract :
Tomotherapy is a special application of intensity modulated radiation therapy based on a geometry similar to computed tomography (CT) imaging. Recently, the viability of using cobalt-60 as the radiation source for tomotherapy has been demonstrated. To further investigate this application an experimental system consisting of a computer controlled first generation translate/rotate CT jig has been constructed. The intensity is modulated by control of the velocity of motion of the phantom through a narrow (1×1 cm2) beam. By comparison, a commercial system developed for linac tomotherapy utilizes the Mimic, a 2D multileaf collimator. The goal of this work is to assess the agreement between an intensity modulated beam generated by these two methods. Beam profiles for a single beam portal are modelled using a pencil-beam based treatment planning system. Beam profiles from a simple wedged beam to more complex profiles are calculated for the multileaf collimator and then an optimal match generated for the scanning pencil beam. Agreement between the profiles is better than 3% in the high dose region. Therefore, beam delivery assessment with our experimental system will be relevant for the investigation of a potential clinical device
Keywords :
computerised tomography; dosimetry; intensity modulation; medical image processing; modelling; radiation therapy; Mimic 2D multileaf collimator; beam delivery assessment; beam profiles; cobalt-60 tomotherapy; computed tomography imaging; computer controlled jig; dosimetric modelling; intensity modulated radiation therapy; linac tomotherapy; optimal match; phantom velocity of motion; scanned pencil beam apparatus; single beam portal; translate/rotate CT jig; treatment planning system; Application software; Biomedical applications of radiation; Collimators; Computational geometry; Computed tomography; Control systems; Geometrical optics; Intensity modulation; Motion control; Optical modulation;
Conference_Titel :
Engineering in Medicine and Biology Society, 2000. Proceedings of the 22nd Annual International Conference of the IEEE
Conference_Location :
Chicago, IL
Print_ISBN :
0-7803-6465-1
DOI :
10.1109/IEMBS.2000.900437