• DocumentCode
    1746678
  • Title

    Measurement of patient exposure dose on X-ray screening mammography

  • Author

    Goto, Sachiko ; Azuma, Yoshiharu ; Sumimoto, Tetsuhiro ; Maruyama, Toshinori ; Eibo, S.

  • Author_Institution
    Fac. of Health Sci., Okayama Univ. Med. Sch., Japan
  • Volume
    1
  • fYear
    2001
  • fDate
    21-23 May 2001
  • Firstpage
    191
  • Abstract
    X-ray mammography is the most effective method to produce a high photographic sensitivity in the early detection of breast cancer in which palpation is impossible. Mammography can be adapted to a screening check aimed at discovering abnormal remarks in females without the subjective symptom. However, in order to justify examining a seemingly healthy group in screening mammography, it is necessary to perform with an appreciation of the risk of carcinogenesis that may be associated with the absorbed radiation dose to the breast (patient exposure dose), and to guarantee an optimization of radiological protection. To do so, it is necessary to control the individual patient exposure dose in screening mammography. Moreover, when estimating patient exposure doses, it is important that breast composition is investigated objectively. In this study, a breast dosimetry system that includes digitization of a mammogram for investigating the individual breast composition was built on the basis of the dosimetry method for quality control of screening mammography. It is thought that this system has 10% or less of an error factor including the error determination of breast composition. Patient exposure doses of 124 examples from 76 patients in Japan were actually estimated by this system. Their values were clearly lower than the maximum acceptable dose recommended by the American College of Radiology and Japan
  • Keywords
    diagnostic radiography; dosimetry; mammography; medical image processing; table lookup; X-ray screening mammography; average glandular dose; breast composition; breast dosimetry system; early breast cancer detection; entrance skin exposure; lookup table; mammogram digitization; mammography QC; patient exposure dose measurement; radiological protection; tissue-equivalent material phantom; Breast cancer; Cancer detection; Dosimetry; Educational institutions; Mammography; Protection; Quality control; Radiology; X-ray detection; X-ray detectors;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Instrumentation and Measurement Technology Conference, 2001. IMTC 2001. Proceedings of the 18th IEEE
  • Conference_Location
    Budapest
  • ISSN
    1091-5281
  • Print_ISBN
    0-7803-6646-8
  • Type

    conf

  • DOI
    10.1109/IMTC.2001.928811
  • Filename
    928811