• DocumentCode
    1752033
  • Title

    Measurement of proportional counter gas gain

  • Author

    Borisov, A. ; Fakhroutdinov, R. ; Goryatchev, V. ; Kojine, A.

  • Author_Institution
    Inst. for High Energy Phys., Protvino, Russia
  • Volume
    1
  • fYear
    2000
  • fDate
    2000
  • Abstract
    A method of evaluation of proportional counter gas gain from arbitrary current measurements has been developed. It provides absolute value of the gain in range from 1 to 108 with moderate current sensitivity and without knowledge of used radioactive source properties. As an example we present dependence of the gain versus voltage for drift tubes of ATLAS Muon Spectrometer filled with mixtures Ar-CO2(93-7) and Ar-N2-CH4(91-4-5) at 3 bar absolute pressure. Results are compared with the gain obtained from ADC measurements
  • Keywords
    argon; muon detection; nitrogen; proportional counters; ADC measurements; ATLAS Muon Spectrometer; Ar-CO2; Ar-CO2(93-7); Ar-N2; Ar-N2-CH4(91-4-5); current measurements; drift tubes; proportional counter gas gain; radioactive source properties; Counting circuits; Current measurement; Detectors; Filling; Gain measurement; Ionization; Mesons; Spectroscopy; Voltage; Wire;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2000 IEEE
  • Conference_Location
    Lyon
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-6503-8
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2000.949094
  • Filename
    949094