Title : 
Development of hydrophones for detecting high-energy reactions in water
         
        
            Author : 
Miyachi, T. ; Goto, S. ; Hasebe, N. ; Higuchi, M. ; Hinode, F. ; Ishiwata, M. ; Kikuchi, R. ; Konno, O. ; Masumura, T. ; Matsuyama, T. ; Misaki, A. ; Nakamura, I. ; Oyamada, M. ; Sato, M. ; Tazawa, Y. ; Tezuka, C.
         
        
            Author_Institution : 
Center for Nucl. Study, Tokyo Univ., Japan
         
        
        
        
        
        
            Abstract : 
Acoustic detectors were developed using a piezo ceramic compound PZT. The shape of the PZT detector was essential to obtain a high sensitivity. A detector of a spherically shaped shell structure, whose size was 50 mm in diameter and 2 mm thick, was fabricated. Its sensitivity was calibrated to be about 40 mV/Pa at 54 kHz. Using the hydrophone, acoustic signals generated by an electron-induced cascade shower in water was detected. Experimental results were compared with simulation data and confirmed consistency
         
        
            Keywords : 
acoustic receivers; hydrophones; piezoelectric devices; PZT; PbZrO3TiO3; acoustic detectors; electron-induced cascade shower; hydrophones; water; Acoustic signal detection; Ceramics; Frequency response; Nuclear and plasma sciences; Physics; Radiation detectors; Shape; Silver; Sonar equipment; Structural shells;
         
        
        
        
            Conference_Titel : 
Nuclear Science Symposium Conference Record, 2000 IEEE
         
        
            Conference_Location : 
Lyon
         
        
        
            Print_ISBN : 
0-7803-6503-8
         
        
        
            DOI : 
10.1109/NSSMIC.2000.949309