Abstract
We report on traceable and spatially resolved intensity measurements of continuous wave (CW) radiation at 100 GHz using laser-based electro-optic (EO) sampling. Traceability of the laser-based setup is achieved by comparison of the EO measurements with measurements using a calibrated radiometer. The uncertainty budget of the EO measurements is discussed in detail. Although we just perform intensity measurements at ~100 GHz, the EO setup can be used for characterization of CW sources over a very broad frequency range from a few gigahertz to ~2.5 THz.
Keywords
calibration; electro-optical devices; intensity measurement; measurement uncertainty; radiometers; sampling methods; terahertz wave imaging; CW GHz radiation; CW sources; EO measurement; THz radiation; absolute intensity measurement; frequency 100 GHz; laser-based electrooptic sampling; measurement uncertainty; radiometer calibration; Antenna measurements; Crystals; Frequency measurement; Measurement by laser beam; Measurement uncertainty; Radiometry; Uncertainty; Absolute intensity measurements; electro-optic (EO) sampling; terahertz (THz) metrology; terahertz (THz) metrology.;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/TIM.2014.2375692
Filename
6983632
Link To Document