Title :
Comment on “Proposal for Measurement Method of Transfer Impedance of Current Probe”
Author :
Lijun Yao ; Weiyu Ma
Author_Institution :
Electromagn. Environ. Effect Lab., Beijing Inst. of Radio Metrol. & Meas., Beijing, China
Abstract :
A previous paper mentioned that the international standard ISO 11452-4:2005 has given a requirement for the calibration fixture of current probe, i.e., the voltage standing wave ratio (VSWR) of the calibration fixture should be less than 1.2. Here is a comment about this.
Keywords :
ISO standards; calibration; electric impedance measurement; measurement standards; VSWR; calibration fixture; current probe; international standard ISO 11452-4:2005; transfer impedance measurement method; voltage standing wave ratio; Calibration; Current measurement; ISO standards; Military standards; Probes; Calibration fixture; ISO 11452-4; calibration jig; current injection probe; current probe; voltage standing wave ratio (VSWR);
Journal_Title :
Electromagnetic Compatibility, IEEE Transactions on
DOI :
10.1109/TEMC.2015.2405612