DocumentCode :
1757911
Title :
Precise Calibration Method for Triaxial Magnetometers Not Requiring Earth’s Field Compensation
Author :
Zikmund, Ales ; Janosek, Michal ; Ulvr, Michal ; Kupec, Josef
Author_Institution :
Dept. of Meas., Czech Tech. Univ. in Prague, Prague, Czech Republic
Volume :
64
Issue :
5
fYear :
2015
fDate :
42125
Firstpage :
1250
Lastpage :
1255
Abstract :
A calibration procedure for calibrations of triaxial magnetometers is presented. The procedure uses a triaxial Helmholtz coil system and an Overhauser scalar magnetometer and is performed in the Earth´s field range. The triaxial coils are first calibrated with the Overhauser magnetometer, and subsequently, a triaxial magnetometer calibration is performed. As opposed to other calibration approaches, neither Earth´s field nulling system nor movements of the magnetometer are needed. A real calibration test was carried out-the extended calibration uncertainty was better than 430 ppm in sensitivity and 0.06° in orthogonality.
Keywords :
calibration; magnetometers; measurement uncertainty; Overhauser scalar magnetometer; extended calibration uncertainty; triaxial Helmholtz coil system; triaxial magnetometers; Calibration; Coils; Current measurement; Earth; Magnetometers; Standards; Uncertainty; Calibration; magnetometer; precision; triaxial; triaxial.;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.2015.2395531