DocumentCode
1759868
Title
A preliminary engineering design of intravascular dual-frequency transducers for contrast-enhanced acoustic angiography and molecular imaging
Author
Jianguo Ma ; Martin, Karl H. ; Dayton, Paul A. ; Xiaoning Jiang
Author_Institution
Dept. of Mech. & Aerosp. Eng., North Carolina State Univ., Raleigh, NC, USA
Volume
61
Issue
5
fYear
2014
fDate
41760
Firstpage
870
Lastpage
880
Abstract
Current intravascular ultrasound (IVUS) probes are not optimized for contrast detection because of their design for high-frequency fundamental-mode imaging. However, data from transcutaneous contrast imaging suggests the possibility of utilizing contrast ultrasound for molecular imaging or vasa vasorum assessment to further elucidate atherosclerotic plaque deposition. This paper presents the design, fabrication, and characterization of a small-aperture (0.6 × 3 mm) IVUS probe optimized for high-frequency contrast imaging. The design utilizes a dual-frequency (6.5 MHz/30 MHz) transducer arrangement for exciting microbubbles at low frequencies (near their resonance) and detecting their broadband harmonics at high frequencies, minimizing detected tissue backscatter. The prototype probe is able to generate nonlinear microbubble response with more than 1.2 MPa of rarefractional pressure (mechanical index: 0.48) at 6.5 MHz, and is also able to detect microbubble response with a broadband receiving element (center frequency: 30 MHz, -6-dB fractional bandwidth: 58.6%). Nonlinear super-harmonics from microbubbles flowing through a 200-μm-diameter micro-tube were clearly detected with a signal-to-noise ratio higher than 12 dB. Preliminary phantom imaging at the fundamental frequency (30 MHz) and dual-frequency super-harmonic imaging results suggest the promise of small aperture, dual-frequency IVUS transducers for contrast-enhanced IVUS imaging.
Keywords
angiocardiography; backscatter; biological tissues; biomechanics; biomedical transducers; biomedical ultrasonics; bubbles; cardiovascular system; minimisation; molecular biophysics; phantoms; atherosclerotic plaque deposition; contrast-enhanced IVUS imaging; contrast-enhanced acoustic angiography; dual-frequency intravascular ultrasound transducers; dual-frequency superharmonic imaging; dual-frequency transducer arrangement; frequency 30 MHz; frequency 6.5 MHz; high-frequency contrast imaging; mechanical index; molecular imaging; nonlinear microbubble response; phantom imaging; rarefractional pressure; signal-to-noise ratio; size 200 mum; tissue backscatter minimization; vasa vasorum assessment; Acoustics; Bandwidth; Imaging; Impedance; Resonant frequency; Transducers; Ultrasonic imaging;
fLanguage
English
Journal_Title
Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
Publisher
ieee
ISSN
0885-3010
Type
jour
DOI
10.1109/TUFFC.2014.2977
Filename
6805699
Link To Document