Abstract :
This paper reports on results obtained from monitoring the respiration and cardiac activity of a patient during a magnetic resonance imaging (MRI) survey using an optical strain sensor based on a fiber Bragg grating. The sensor is proposed specifically to acquire ballistocardiographic signals from a patient exposed to high intensity electromagnetic radiation. A Bland-Altman analysis shows the measurements that have a satisfactory accuracy for monitoring purposes, and the relative error is . The method is both noninvasive and safe for the patient. In addition, the sensor does not affect the MRI imaging quality.
Keywords :
Bragg gratings; biomedical MRI; biomedical equipment; cardiology; fibre optic sensors; patient monitoring; pneumodynamics; strain sensors; Bland-Altman analysis; MRI imaging; ballistocardiographic signals; cardiac activity monitoring; fiber Bragg grating strain sensor; high intensity electromagnetic radiation exposure; magnetic resonance imaging survey; optical strain sensor; patient vital signs; respiration monitoring; Fiber gratings; Heart rate; Magnetic resonance imaging; Monitoring; Optical fiber sensors; Strain; Ballistocardiography (BCG); fiber Bragg gratings (FBGs); heart rate (HR); magnetic resonance imaging (MRI); respiration rate (RR); strain sensors;
Journal_Title :
Sensors Journal, IEEE
DOI :
10.1109/JSEN.2013.2279160