DocumentCode
1762268
Title
Instantaneous microwave imaging with time-domain measurements for breast cancer detection
Author
Kwon, Susanna ; Lee, Sang-Rim
Author_Institution
Dept. of Electr. Eng., Ewha Womans Univ., Seoul, South Korea
Volume
49
Issue
10
fYear
2013
fDate
May 9 2013
Firstpage
639
Lastpage
641
Abstract
A breast cancer imaging system with time-domain measurements is proposed. This system is based on complementary metal-oxide semiconductor circuits such that it is very cost-effective and the image is acquired in a very short time. It consists of ultra-wideband transceivers and a master controller for system control and data collection. Simulation results with a 3D FDTD hemi-spherical breast model show the proposed system can reconstruct the breast image in the presence of possible disturbances such as clock skew, clock jitter, and white noise.
Keywords
CMOS image sensors; biomedical electronics; biomedical equipment; image denoising; image reconstruction; mammography; medical control systems; medical image processing; microwave imaging; phantoms; physiological models; 3D FDTD hemispherical breast model; CMOS circuits; breast cancer detection; breast cancer imaging system; breast image reconstruction; clock jitter; clock skew; complementary metal oxide semiconductor; data collection; instantaneous microwave imaging; master controller; system control; time domain measurements; ultra wideband transceivers; white noise;
fLanguage
English