Title :
Electromagnetic Thermotherapy for Deep Organ Ablation by Using a Needle Array Under a Synchronized-Coil System
Author :
Sheng-Chieh Huang ; Jui-Wen Kang ; Hung-Wen Tsai ; Yan-Shen Shan ; Xi-Zhang Lin ; Gwo-Bin Lee
Author_Institution :
Dept. of Power Mech. Eng., Nat. Tsing Hua Univ., Hsinchu, Taiwan
Abstract :
Thermal ablation by using electromagnetic thermotherapy (EMT) has been a promising cancer modality in recent years. It has relatively few side effects and has therefore been extensively investigated for a variety of medical applications in internal medicine and surgery. The EMT system applies a high-frequency alternating electromagnetic field to heat up the needles which are inserted into the target tumor to cause tumor ablation. In this study, a new synchronized-coil EMT system was demonstrated, which was equipped with two synchronized coils and magnetic field generators to provide a long-range, penetrated electromagnetic field to effectively heat up the needles. The heating effect of the needles at the center of the two coils was first explored. The newly designed two-section needle array combined with the synchronized-coil EMT system was thus demonstrated in the in vitro and in vivo animal experiments. Experimental data showed that the developed system is promising for minimally invasive surgery since it might provide superior performance for thermotherapy in cancer treatment.
Keywords :
biothermics; cancer; electromagnetic fields; surgery; tumours; cancer modality; cancer treatment; deep ablation; electromagnetic field; electromagnetic thermotherapy; heating effect; in vitro animal experiments; in vivo animal experiments; magnetic field generators; minimally invasive surgery; needle array; synchronized-coil EMT system; synchronized-coil system; target tumor ablation; thermal ablation; thermotherapy; two-section needle array; Coils; Electromagnetic fields; Heating; Magnetic cores; Needles; Synchronization; Temperature measurement; Ablation; electromagnetic thermotherapy (EMT) system; minimally invasive surgery (MIS); needle arrays; synchronized-coil system;
Journal_Title :
Biomedical Engineering, IEEE Transactions on
DOI :
10.1109/TBME.2014.2339499